Site visit or send-in evaluation
Send the readhead, ring and interface labels, ring diameter and section, controller model, mounting photos and when the alarm occurs. Tell us if the alarm repeats in the same angular region. Use the enquiry form to share the details.
What specialist testing can show
Use the Renishaw setup LEDs and compatible Siemens diagnostics to check the drive link, head communication and signal quality around the complete ring. A dial indicator checks runout separately from the optical readout. Safety-related commissioning requires the complete validated procedure for the configured axis. Exact Mitchell TI-5000JX and TI-3000JX support for this optical DRIVE-CLiQ system was not verified. Renishaw signal and safety verification (PDF)
Renishaw RESOLUTE RESA30 DRIVE-CLiQ FS Encoder technical reference
Find your part number
Read the part numbers of all three items. The readhead is ordered for a ring diameter and interface. The A-section and B-section RESA30 rings have separate mounting drawings. The single-input and dual-input interface boxes also differ.
| Order identifier | Part | What to match |
|---|---|---|
| SA29DBA052B30S | Renishaw FS angular-head order example | 29-bit Siemens DRIVE-CLiQ, Ø52 mm ring size, standard outlet, 3 m lead and sealed M12. |
| SA26D… | 26-bit FS angular head family | Read the entire code for diameter, outlet, cable and termination. |
| A-9796-0575 | Single-head DRIVE-CLiQ interface | Check revision and approved head and ring pair. |
| A-9796-0590 | Dual-head DRIVE-CLiQ interface | Use the approved dual-head mechanical and commissioning setup. |
The 052 field describes ring diameter. It does not identify the ring’s A/B cross-section or the required hub by itself. Renishaw FS order nomenclature (PDF) · RESA30 sections and sizes (PDF)
Production and end-of-life status
Renishaw provides current RESOLUTE FS and RESA30 documentation. Confirm orderability and matching replacement for the exact full head, ring and interface numbers. Reviewed 7 November 2026. Renishaw RESOLUTE series
Specifications
| Parameter | Value / condition |
|---|---|
| Measurement | Optical absolute singleturn readhead with a stainless-steel RESA30 ring. Nominal scale pitch 30 µm. |
| Interface / resolution | Siemens DRIVE-CLiQ via separate interface. SA29D order example uses 29-bit position. Installed accuracy depends on ring and mounting. |
| Supply | 24 V from the DRIVE-CLiQ network to the interface. The sealed readhead connection has a separate 5 V circuit. System power ≤4.3 W in the safety manual. |
| Head geometry | Nominal 0.8 mm rideheight, ±0.1 mm for a 52 mm ring in the referenced drawing. Standard rotary drawing gives yaw, pitch and roll tolerances of ±0.5°. |
| Environment | Standard head operating 0 to +80 °C and IP64. Interface operating 0 to +55 °C and IP67 with specified connections. |
Resolution is the size of the reported increment. Ring eccentricity, tilt and mounting distortion still affect position accuracy. Renishaw rotary installation and system specifications (PDF) · RESA30 installed accuracy (PDF)
Mounting and fit
Mount the ring to the correct A- or B-section hub, then check its seating and runout with the method in the relevant drawing. Keep the scale surface clean. Use the readhead mounting drawing to set rideheight, pitch, yaw and roll. The Ø52 mm ring has a tighter rideheight allowance than the general drawing.
For the standard FS readhead, the rotary drawing specifies M3 mounting and recommended tightening torque 0.9–1.1 N·m. This value is for the standard readhead, separate from the ring fasteners and the UHV head. Check signal quality around the full ring. An excellent signal at one point can hide a runout or alignment problem elsewhere. Renishaw rotary head and ring installation drawings (PDF) · RESA30 A/B mechanical drawings (PDF)
Interface and measuring method
The readhead reads the absolute pattern on the rotating ring and sends digital position through its link to the interface box. The interface connects to the Siemens DRIVE-CLiQ system, which carries communication and interface power. The two links have different signal and supply assignments. A controller can see the interface even if the head-to-ring optical signal is weak.
Verified wiring
| Circuit | DRIVE-CLiQ link at interface | Sealed readhead M12, termination S |
|---|---|---|
| Supply / return | 24 V pin 1, 0 V pin 5 | 5 V pin 2, 0 V pins 5 and 8 |
| Communications | RX+ / RX− pins 3 / 4. TX+ / TX− pins 7 / 6. | A+ / A− pins 3 / 4 |
| Shield and reserved contacts | Case shield. Use specified network lead. | Case shield. Pins 6 and 7 reserved. |
Use the connector-face drawings and the interface port labels. The table is for the named Renishaw system. Apply the 24 V network supply only to the network link, and use the approved readhead-to-interface assembly. Use the dedicated JST map for an F or V termination. Renishaw output signals and connector drawings (PDF)
Cables inside the motor and at the machine
| Part or code | Application |
|---|---|
| SA29DBA052B30S | Reference readhead with integral 3 m cable and sealed M12 termination. |
| A-9796-0575 / A-9796-0590 | Single- or dual-input interface module. These are interface part numbers. |
| DRIVE-CLiQ cable | Separate approved cable from the Renishaw interface to the exact Siemens system. Confirm both ends, length and machine cable ID. |
The RESA30 is an external axis feedback system. It has no universal in-motor loom. Use the maker’s approved assemblies for safety-related service, with the original routing and shield terminations. Readhead cable codes and interface drawings (PDF) · Approved installation and electrical requirements (PDF)
Fault checks
Secure the axis and isolate it before checking the ring or connectors. Follow your site’s lockout procedure. Secure the load against movement, isolate all motor, drive and separate feedback supplies, allow the specified discharge time and verify absence of voltage. A qualified technician should perform internal and electrical checks. Unplug or reconnect only with the applicable circuits safely de-energized.
- Inspect. Check the connector key and latch, bent pins, deposits, damaged shielding and cable support. Look for a displaced ring, rubbing, contamination or loose head bracket.
- Check a separate detachable cable. With the harness isolated from the encoder and receiver at both ends, use resistance mode. First short the meter probes to establish lead resistance. Check each conductor end to end and flex the cable gently at the strain relief. Check between unrelated wires and from signal wires to shield or shell. Expect open only where the drawing calls for isolation. Compare any intentional sense links or shield bonds with the equipment drawing. Keep insulation testers away from connected electronics. Leave the integral readhead and interface electronics out of a resistance test.
- Check each link. With the OEM-approved breakout fitted while off, a qualified person can verify the network supply at the interface from pin 1 to 5. Use approved diagnostics for head power and communication. The two M12 connectors have different supplies.
- Interpret position-dependent faults. A stable link and intermittent head signal at the same angle point toward ring runout, head geometry or scale condition. No interface response calls for network power and cable checks. Good position with a safety alarm calls for verifying the validated device combination and safety commissioning.
Motor and drive applications
This guide covers the Renishaw FS rotary readhead and RESA30 ring with the dedicated Siemens DRIVE-CLiQ interface. Identify the exact Siemens control or drive, firmware, and safety configuration before replacement. Other RESOLUTE serial families and REXA30 rings have separate configuration rules. Renishaw approved system combinations (PDF)
Angle of alignment and commissioning
Set mechanical alignment and check optical signal quality around a full turn before establishing the axis datum. Record the incoming scaling, direction and machine zero. Complete the Siemens and Renishaw commissioning and applicable safety checks for the actual axis. A motor commutation angle applies only when the system is configured for that purpose by the machine maker.
Drawings, pin maps and manuals
- Renishaw FS Siemens DRIVE-CLiQ installation and safety manual (PDF). Rotary readhead setup, RESA30 A/B ring drawings, safety commissioning and the two M12 pinouts.
- Renishaw RESOLUTE FS data sheet (PDF). Use the Siemens DRIVE-CLiQ angular order table and interface diagrams.
- Renishaw RESA30 ring data sheet (PDF). Ring section, diameter, fit and accuracy.
- Renishaw RESOLUTE downloads. Use the manual edition for the exact hardware.